FAQ
Q: Is the pump open system?
A: Yes, Universal IV set can be used with our Infusion Pump after calibration.
Q: Does the pump compatible with Micro IV Set (1 ml=60 drops)?
A: Yes, all of our pumps are compatible with IV Set of 15/20/60 dorps.
Q: Is it a peristaltic pumping mechanism?
A: Yes, curvilinear peristaltic.
Q: What is the difference between PURGE and BOLUS function?
A: Purge used to remove air in line before infusion. Bolus can be administered for infusion therapy during infusion. Both purge and bolus rate are programmable.
Specifications
Model |
ZNB-XAII |
Pumping Mechanism |
Curvilinear peristaltic |
IV Set |
Compatible with IV sets of any standard |
Flow Rate |
1-1500 ml/h (in 0.1 ml/h increments) |
Purge, Bolus |
100-1500 ml/h (in 0.1 ml/h increments)
Purge when pump stops, bolus when pump starts |
Accuracy |
±3% |
*Inbuilt Thermostat |
30-45℃, adjustable |
VTBI |
1-20000 ml (in 0.1 ml increments) |
Infusion Mode |
ml/h, drop/min, time-based, body weight, nutrition |
KVO Rate |
0.1-5 ml/h (in 0.1 ml/h increments) |
Alarms |
Occlusion, air-in-line, door open, end program, low battery, end battery,
AC power off, motor malfunction, system malfunction, standby |
Additional Features |
Real-time infused volume, automatic power switching, mute key,
purge, bolus, system memory, history log, key locker, drug library,
rotary knob, change flow rate without stopping the pump |
Drug Library |
Available |
Occlusion Sensitivity |
High, medium, low |
History Log |
50000 events |
Air-in-line Detection |
Ultrasonic detector |
Wireless management |
Optional |
Drop Sensor |
Optional |
Vehicle Power (Ambulance) |
12±1.2 V |
Power Supply, AC |
110/230 V (optional), 50-60 Hz, 20 VA |
Battery |
9.6±1.6 V, rechargeable |
Battery Life |
5 hours at 25 ml/h |
Working Temperature |
10-30℃ |
Relative Humidity |
30-75% |
Atmospheric Pressure |
860-1060 hpa |
Size |
130*145*228 mm |
Weight |
2.5 kg |
Safety Classification |
Class Ⅰ, type CF |