Home Categories Send inquiry

KL-605T Syringe Pump

FAQ Q: Automatic recognition of syringe size and fixation A: Yes. Q: Syringe barrel clamp alarm? A: Yes, it is Syringe Error alarm. Q: Syringe plunger disengaged alarm? A: Yes, it is Installation Error alarm. Q: Automatic anti-bolus? A: Yes, anti-bolus system to reduce pressure on sudden release of occlusion. Q: Is it capable of horizontal stacking of more than two pumps? A: Yes, it is stackable up to 4 pumps or 6 pumps.   Specifications
Model KL-605T
Syringe Size 5, 10, 20, 30, 50/60 ml
Applicable Syringe Compatible with syringe of any standard
VTBI 1-1000 ml (in 0.1, 1, 10 ml increments)
Flow Rate Syringe 5 ml: 0.1-100 ml/h (in 0.01, 0.1, 1, 10 ml/h increments) Syringe 10 ml: 0.1-300 ml/h Syringe 20 ml: 0.1-600 ml/h Syringe 30 ml: 0.1-800 ml/h Syringe 50/60 ml: 0.1-1200 ml/h
Bolus Rate Same as flow rate
Anti-Bolus Automatic
Accuracy ±2% (mechanical accuracy≤1%)
Infusion Mode Flow rate: ml/min, ml/h Time-based Body weight: mg/kg/min, mg/kg/h, ug/kg/min, ug/kg/h etc.
KVO Rate 0.1-1 ml/h (in 0.01 ml/h increments)
Alarms Occlusion, near empty, end program, low battery, end battery, AC power off, motor malfunction, system malfunction, standby, pressure sensor error, syringe installation error, syringe drop off
Additional Features Real-time infused volume, automatic power switching, automatic syringe identification, mute key, purge, bolus, anti-bolus, system memory, history log, key locker
Drug Library Available
Occlusion Sensitivity High, medium, low
Docking Station Stackable up to 4-in-1 or 6-in-1 Docking Station with single power cord
History Log 50000 events
Wireless management Optional
Power Supply, AC 110/230 V, 50/60 Hz, 20 VA
Battery 14.8 V, rechargeable
Battery Life 8 hours at 5 ml/h
Working Temperature 5-40℃
Relative Humidity 20-90%
Atmospheric Pressure 700-1060 hpa
Size 245*120*115 mm
Weight 2.5 kg
Safety Classification Class Ⅱ, type BF

KL-605T Syringe Pump Related Video: