FAQ
Q: Automatic recognition of syringe size and fixation?
A: Yes.
Q: Syringe barrel clamp alarm?
A: Yes, it is Syringe Error alarm.
Q: Syringe plunger disengaged alarm?
A: Yes, it is Installation Error alarm.
Q: Automatic anti-bolus?
A: Yes, anti-bolus system to reduce pressure on sudden release of occlusion.
Q: Is it capable of horizontal stacking of more than two pumps?
A: Yes, it is stackable up to 4 pumps or 6 pumps.
Specifications
| Model |
KL-605T |
| Syringe Size |
5, 10, 20, 30, 50/60 ml |
| Applicable Syringe |
Compatible with syringe of any standard |
| VTBI |
1-1000 ml (in 0.1, 1, 10 ml increments) |
| Flow Rate |
Syringe 5 ml: 0.1-100 ml/h (in 0.01, 0.1, 1, 10 ml/h increments)
Syringe 10 ml: 0.1-300 ml/h
Syringe 20 ml: 0.1-600 ml/h
Syringe 30 ml: 0.1-800 ml/h
Syringe 50/60 ml: 0.1-1200 ml/h |
| Bolus Rate |
Same as flow rate |
| Anti-Bolus |
Automatic |
| Accuracy |
±2% (mechanical accuracy≤1%) |
| Infusion Mode |
Flow rate: ml/min, ml/h
Time-based
Body weight: mg/kg/min, mg/kg/h, ug/kg/min, ug/kg/h etc. |
| KVO Rate |
0.1-1 ml/h (in 0.01 ml/h increments) |
| Alarms |
Occlusion, near empty, end program, low battery, end battery,
AC power off, motor malfunction, system malfunction, standby,
pressure sensor error, syringe installation error, syringe drop off |
| Additional Features |
Real-time infused volume, automatic power switching,
automatic syringe identification, mute key, purge, bolus, anti-bolus,
system memory, history log, key locker |
| Drug Library |
Available |
| Occlusion Sensitivity |
High, medium, low |
| Docking Station |
Stackable up to 4-in-1 or 6-in-1 Docking Station with single power cord |
| History Log |
50000 events |
| Wireless management |
Optional |
| Power Supply, AC |
110/230 V, 50/60 Hz, 20 VA |
| Battery |
14.8 V, rechargeable |
| Battery Life |
8 hours at 5 ml/h |
| Working Temperature |
5-40℃ |
| Relative Humidity |
20-90% |
| Atmospheric Pressure |
700-1060 hpa |
| Size |
245*120*115 mm |
| Weight |
2.5 kg |
| Safety Classification |
Class Ⅱ, type BF |